3500 Genetic Analyzer for Resequencing & Fragment Analysis

Specifications
General Specifications
| Platform: |
3500 Genetic Analyzer
|
| Format: |
96-well plate,
8-well strips
|
| Weight: |
82 kg (approximately)
|
| Throughput: |
Up to 424 samples per day (using run module HID36 POP-7)
|
| Read Length: |
≥850 bp using a 50cm capillary and POP-7
|
| Product Size: |
1 system
|
| Current Limit: |
15 amps
|
| Plan Duration: |
Service Installation and Application Training Included,
1 Year Limited Warranty on Parts and Labor
|
| Voltage Limit: |
Electrophoresis Voltage - Up to 20 kV
|
| Capillary Length: |
50 cm,
36 cm
|
| Operating System: |
Windows Vista (Service Pack 1)
|
| Analysis Software: |
GeneMapper® Software
|
| External Computer: |
Operating System: Windows® Vista® SP1,
Installed RAM: 2GB,
Hard Drive: 1 x 80GB 7200 RPM SATA 3.0Gb⁄s and 8MB Data Burst Cache,
Hardware: Pentium® IV Processor - 1.86 GHz Processor
|
| Compatible Polymers: |
POP-7,
POP-6,
POP-4
|
| Compatible Products: |
3500 Genetic Analyzer Capillary Array (50 cm),
3500 Genetic Analyzer Capillary Array (36 cm),
POP-6™ Polymer for 3500⁄3500xL Genetic Analyzers,
POP-7™ Polymer for 3500⁄3500xL Genetic Analyzers,
POP-4™ Polymer for 3500⁄3500xL Genetic Analyzers
|
| Polymer Consumption: |
Replace at earliest of 7 days⁄384 samples⁄60 Injections (384-sample pouches),
Replace at earliest of 7 days⁄960 samples⁄120 Injections (960-sample pouches)
|
| Regulatory Statement: |
For Research Use Only. Not for use in diagnostic procedures.
|
| Number of Capillaries: |
8 Capillaries
|
| Operating Environment: |
Room temp: <2° fluctuation during runs,
Humidity: 20%-80% (non-condensing),
Temperature: 15°C-30°C
|